FDA Enforcement Class II Ongoing

LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.

Recall: Z-0229-2026 · Reported October 29, 2025

Enforcement

Recall Number
Z-0229-2026
Event ID
97693
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Synthes (USA) Products LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
October 29, 2025
Initiation Date
September 26, 2025
Classification Date
October 21, 2025
Address
1301 Goshen Pkwy, West Chester, PA, 19380-5986, United States

Description

LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.

Reason

Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

Code Info

Product Code: 04.045.342TS. GTIN: 10886982301390. Lot Number: 31756P9. Expiration Date: 08/31/2025

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.

Quantity

122 units