Synchro2, 0.014in x 200cm STANDARD Guidewire with Hydrophilic Coating, UPN Product No. M00326410, REF 2641, STERILE: EO - Product Usage: product is intended for general intravascular use, including neurovascular and peripheral vasculatures.
Enforcement
- Recall Number
- Z-0229-2020
- Event ID
- 81108
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Neurovascular
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 6, 2019
- Initiation Date
- June 8, 2018
- Classification Date
- October 29, 2019
- Termination Date
- April 28, 2021
- Address
- 47900 Bayside Pkwy, Fremont, CA, 94538-6515, United States
Description
Synchro2, 0.014in x 200cm STANDARD Guidewire with Hydrophilic Coating, UPN Product No. M00326410, REF 2641, STERILE: EO - Product Usage: product is intended for general intravascular use, including neurovascular and peripheral vasculatures.
The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.
UPN: M00326410 Lot Numbers/UDI:0000000564/ (01)04546540688767(17)201031(10) 0000000564; 0000000565/ (01)04546540688767(17)201031(10)0000000565; 0000000566/(01)04546540688767(17)201031(10)0000000566; 0000000673/ (01)04546540688767(17)201031(10)0000000673; 0000000674/ (01)04546540688767(17)201031(10)0000000674; 0000000675/ (01)04546540688767(17)201031(10)0000000675; 0000000983/ (01)04546540688767(17)201031(10)0000000983; 0000000982/ (01)04546540688767(17)201031(10)0000000982; 0000001016/ (01)04546540688767(17)201031(10)0000001016; 0000001017/ (01)04546540688767(17)201031(10)0000001017; 0000001018/ (01)04546540688767(17)201031(10)0000001018; 0000001280/ (01)04546540688767(17)201031(10)0000001280; 0000001279/ (01)04546540688767(17)201031(10)0000001279; 0000001278/ (01)04546540688767(17)201031(10)0000001278; 0000001289/ (01)04546540688767(17)201031(10)0000001289; 0000001290/ (01)04546540688767(17)201031(10)0000001290; 0000001291/ (01)04546540688767(17)201031(10)0000001291; 0000001495/ (01)04546540688767(17)201031(10)0000001495; 0000001497/ (01)04546540688767(17)201031(10)0000001497; 0000001496/ (01)04546540688767(17)201031(10)0000001496; 0000001640/ (01)04546540688767(17)201130(10)0000001640; 0000001641/ (01)04546540688767(17)201130(10)0000001641; 0000001642/ (01)04546540688767(17)201130(10)0000001642; 0000000989/ (01)04546540688767(17)201031(10)0000000989; 0000001912/ (01)04546540688767(17)201130(10)0000001912; 0000001913/ (01)04546540688767(17)201130(10)0000001913; 0000001914/ (01)04546540688767(17)201130(10)0000001914; 0000001643/ (01)04546540688767(17)201130(10)0000001643; 0000002056/ (01)04546540688767(17)201130(10)0000002056; 0000002057/ (01)04546540688767(17)201130(10)0000002057; 0000002058/ (01)04546540688767(17)201130(10)0000002058; 0000002236/ (01)04546540688767(17)201130(10)0000002236; 0000002235/ (01)04546540688767(17)201130(10)0000002235; 0000002233/ (01)04546540688767(17)201130(10)0000002233; 0000002234/ (01)04546540688767(17)201130(10)0000002234; 0000002427/ (01)04546540688767(17)201231(10)0000002427; 0000002428/ (01)04546540688767(17)201231(10)0000002428; 0000002429/ (01)04546540688767(17)201231(10)0000002429; 0000002430/ (01)04546540688767(17)201231(10)0000002430; 0000002642/ (01)04546540688767(17)201231(10)0000002642; 0000002641/ (01)04546540688767(17)201231(10)0000002641; 0000002643/ (01)04546540688767(17)201231(10)0000002643; 0000002644/ (01)04546540688767(17)201231(10)0000002644; 0000002766/ (01)04546540688767(17)201231(10)0000002766; 0000002767/ (01)04546540688767(17)201231(10)0000002767; 0000002768/ (01)04546540688767(17)201231(10)0000002768; 0000002974/ (01)04546540688767(17)201231(10)0000002974; 0000002975/ (01)04546540688767(17)201231(10)0000002975; 0000002976/ (01)04546540688767(17)201231(10)0000002976; 0000003768/ (01)04546540688767(17)210131(10)0000003768; 0000003767/ (01)04546540688767(17)210131(10)0000003767; 0000003769/ (01)04546540688767(17)210131(10)0000003769; 0000003770/ (01)04546540688767(17)210131(10)0000003770; 0000003950/ (01)04546540688767(17)210131(10)0000003950; 0000004126/ (01)04546540688767(17)210131(10)0000004126; 0000004125/ (01)04546540688767(17)210131(10)0000004125; 0000004111/ (01)04546540688767(17)200124(10)0000004111; 0000004321/ (01)04546540688767(17)200130(10)0000004321; 0000004597/ (01)04546540688767(17)200205(10)0000004597; 0000004899/ (01)04546540688767(17)210228(10)0000004899; 0000004900/ (01)04546540688767(17)210228(10)0000004900; 0000004901/ (01)04546540688767(17)210228(10)0000004901; 0000005142/ (01)04546540688767(17)210228(10)0000005142; 0000005144/ (01)04546540688767(17)210228(10)0000005144; 0000005145/ (01)04546540688767(17)210228(10)0000005145; 0000005146/ (01)04546540688767(17)210228(10)0000005146; 0000005148/ (01)04546540688767(17)210228(10)0000005148; 0000005344/ (01)04546540688767(17)210228(10)0000005344; 0000005346/ (01)04546540688767(17)210228(10)0000005346; 0000005345/ (01)04546540688767(17)210228(10)0000005345; 0000005347/ (01)04546540688767(17)210228(10)0000005347; 0000005348/ (01)04546540688767(17)210228(10)0000005348; 0000005532/ (01)04546540688767(17)210331(10)0000005532; 0000005533/ (01)04546540688767(17)210331(10)0000005533; 0000005531/ (01)04546540688767(17)200301(10)0000005531; 0000005687/ (01)04546540688767(17)200305(10)0000005687; 0000005828/ (01)04546540688767(17)200307(10)0000005828; 0000004902/ (01)04546540688767(17)210228(10)0000004902; 0000004903/ (01)04546540688767(17)210228(10)0000004903; 0000005852/ (01)04546540688767(17)210331(10)0000005852; 0000005853/ (01)04546540688767(17)210331(10)0000005853; 0000005987/ (01)04546540688767(17)210331(10)0000005987; 0000005988/ (01)04546540688767(17)210331(10)0000005988; 0000005989/ (01)04546540688767(17)210331(10)0000005989; 0000006199/ (01)04546540688767(17)210331(10)0000006199; 0000006200/ (01)04546540688767(17)210331(10)0000006200; 0000006316/ (01)04546540688767(17)210331(10)0000006316; 0000006315/ (01)04546540688767(17)210331(10)0000006315; 0000006314/ (01)04546540688767(17)210331(10)0000006314; 0000006601/ (01)04546540688767(17)210325(10)0000006601; 0000006672/ (01)04546540688767(17)210326(10)0000006672; 0000006482/ (01)04546540688767(17)210331(10)0000006482; 0000006480/ (01)04546540688767(17)210331(10)0000006480; 0000006481/ (01)04546540688767(17)210331(10)0000006481; 0000006827/ (01)04546540688767(17)210401(10)0000006827; 0000006619/ (01)04546540688767(17)210331(10)0000006619; 0000006618/ (01)04546540688767(17)210331(10)0000006618; 0000006620/ (01)04546540688767(17)210331(10)0000006620; 0000007032/ (01)04546540688767(17)210407(10)0000007032; 0000007001/ (01)04546540688767(17)210430(10)0000007001; 0000006999/ (01)04546540688767(17)210430(10)0000006999; 0000007000/ (01)04546540688767(17)210430(10)0000007000; 0000007174/ (01)04546540688767(17)210410(10)0000007174; 0000007082/ (01)04546540688767(17)210408(10)0000007082; 0000007089/ (01)04546540688767(17)210430(10)0000007089; 0000007090/ (01)04546540688767(17)210430(10)0000007090; 0000007091/ (01)04546540688767(17)210430(10)0000007091; 0000007092/ (01)04546540688767(17)210410(10)0000007092; 0000007315/ (01)04546540688767(17)210430(10)0000007315; 0000007317/ (01)04546540688767(17)210430(10)0000007317; 0000007316/ (01)04546540688767(17)210430(10)0000007316; 0000007426/ (01)04546540688767(17)210430(10)0000007426; 0000007425/ (01)04546540688767(17)210430(10)0000007425; 0000007741/ (01)04546540688767(17)210422(10)0000007741; 0000007771/ (01)04546540688767(17)210423(10)0000007771; 0000007772/ (01)04546540688767(17)210423(10)0000007772; 0000009017.
Worldwide distributions- U.S. Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, DE, FL, GA, GU, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY. Countries of Austria, Argentina, Australia, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Lithuania, Mexico, Netherlands, Norway, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, UAE, and UK.
21971