FDA Enforcement Class II Terminated

BACT/ALERT VIRTUO system, A Unit, with Version R2.0 Firmware Product Usage: The BacT/ALERT VIRTU Microbial Detection System is an automated microbial test system capable of incubating, agitating, and continuously monitoring for the detection of aerobic, facultative, and anaerobic microorganism growth from blood and other normally sterile body fluids.

Recall: Z-0228-2019 · Reported October 31, 2018

Enforcement

Recall Number
Z-0228-2019
Event ID
80759
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
bioMerieux, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 31, 2018
Initiation Date
June 27, 2018
Classification Date
October 23, 2018
Termination Date
July 21, 2020
Address
100 Rodolphe St, N/A, Durham, NC, 27712-9402, United States

Description

BACT/ALERT VIRTUO system, A Unit, with Version R2.0 Firmware Product Usage: The BacT/ALERT VIRTU Microbial Detection System is an automated microbial test system capable of incubating, agitating, and continuously monitoring for the detection of aerobic, facultative, and anaerobic microorganism growth from blood and other normally sterile body fluids.

Reason

The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells P24 - P27, where the calibration standards typically reside. If the calibration standards are not returned to the designated locations, an anomaly may occur.

Code Info

Reference Number 411660, UDI 03573026369767, all serial numbers

Distribution

Worldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KS, LA, MI, MO, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SD, TN, TX, UT, VT, WV. Distributed internationally to Austria, Australia, Belgium, Brazil, Canada, Switzerland, Chile, Colombia, Germany, Spain, France, United Kingdom, Hong Kong, India, Italy, Japan, South Korea, Netherlands, Portugal, Sweden, Singapore, Thailand, South Africa, Saudi Arabia, Philippines, Myanmar, Serbia, Lebanon, United Arab Emirates, Slovenia, Qatar, Costa Rica, Croatia, Panama, Estonia, Israel, Philippines, Malaysia, Brunei Darussalam, Vietnam, Kuwait.

Quantity

516 units