FDA Enforcement Class I Terminated

Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System. Intended for posterior pedicle screw fixation (Tl OS2- ilium), posterior hook fixation (Tl-L5), or anterolateral fixation (T8-L5).

Recall: Z-0226-2015 · Reported December 3, 2014

Enforcement

Recall Number
Z-0226-2015
Event ID
69603
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Synthes (USA) Products LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 3, 2014
Initiation Date
October 23, 2014
Classification Date
November 26, 2014
Termination Date
August 10, 2015
Address
1301 Goshen Pkwy, N/A, West Chester, PA, 19380-5986, United States

Description

Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System. Intended for posterior pedicle screw fixation (Tl OS2- ilium), posterior hook fixation (Tl-L5), or anterolateral fixation (T8-L5).

Reason

Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct specification.

Code Info

Part Number 04.634.002 Lot Number 6816781

Distribution

Distributed to MI, MT, and CO.

Quantity

24