FDA Enforcement Class II Terminated

Micro Surgery Handpiece SGA, Model SGA-ES, Order Code:H263, Manufactured by: Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

Recall: Z-0225-2019 · Reported October 31, 2018

Enforcement

Recall Number
Z-0225-2019
Event ID
80713
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nakanishi Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 31, 2018
Initiation Date
June 1, 2018
Classification Date
October 23, 2018
Termination Date
April 28, 2021
Address
700, Shimohinata, Kanuma, N/A, N/A, Japan

Description

Micro Surgery Handpiece SGA, Model SGA-ES, Order Code:H263, Manufactured by: Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

Reason

Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

Code Info

Previous Operation Manuals: OM-SH0015E002, OM-SH0015E003, OM-SH0015E004, OM-SH0015E005, OM-SH0100F, OM-SH0096S. Model Name: SGA-ES Order Code: H263 0BB10068 - 0BB10069; 0BB20006 -0BB20010; 0BB30044-0BB30046, 0BB30051 - 0BB30053; 0BB60035 - 0BB60042; 0BB70021, 0BB70022, 0BB70025, 0BB70026, 0BB60063 - 0BB60072; 0BB70020, 0BB70023, 0BB70024,0BBX0042 - 0BBX0044; 0BC10046 - 0BC10070; 0BC20053 - 0BC20060; 0BC30011 - 0BC30014, 0BC30081 -0BC30087; 0BC40041- 0BC40043; 0BC40054 -0BC40063; 0BC50009- 0BC50018; 0BC50041-0BC50044, 0BC50046- 0BC50055, 0BC50065, 0BC50066, 0BC50073, 0BC50074, 0BC60010- 0BC60019; 0BC90031 - 0BC90035; 0BCY0086 - 0BCY0095; 0BD50003 -0BD50012; ABDZ0011 -ABDZ0020; ABDZ0083-ABDZ0102, ABDZ0112-ABDZ0117; ABE30001-ABE30015 , ABE30017- ABE30034; ABE40069 - ABE40088, ABE40095, ABE40096, ABE50001- ABE50010, ABE50012- ABE50021; ABE40090; ABE50036 - ABE50040; ABE90009 - ABE90018; ABE90041 - ABE90050; ABEY0001- ABEY0010; ABF40022 - ABF40041; ABF50031- ABF50040; ABF70020 - ABF70025; ABF70058 - ABF70065; ABFY0021- ABFY0030, ABFX0030, ABFY0001 - ABFY0006, ABFY0008-ABFY0010; ABFZ0034- ABFZ0039, ABFZ0041, ABFZ0042, ABFZ0044- ABFZ0053;ABG40011-ABG40022; ABG40024; ABG70128, ABG70129, ABG70131 - ABG70140; ABG80001-ABG80018; ABG90125 -ABG90127, ABG90129, ABG90130, ABG90132, ABG90133; ABH30023.

Distribution

U.S. Distribution to states of: CA, FL, GA, IL, WA.; and internationally to: Canada.

Quantity

440