FDA Enforcement Class II Ongoing

NEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540

Recall: Z-0223-2023 · Reported November 16, 2022

Enforcement

Recall Number
Z-0223-2023
Event ID
91047
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
CooperSurgical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 16, 2022
Initiation Date
October 20, 2022
Classification Date
November 10, 2022
Address
95 Corporate Dr, N/A, Trumbull, CT, 06611-1350, United States

Description

NEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540

Reason

Firm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In one case, the patient suffered a laceration to the throat.

Code Info

Lot/Serial Number UDI-DI (GTIN-14 - not yet added to GUDID database) 272992 00888937016659 269476 00888937016659 269477 00888937016659 269478 00888937016659 269479 00888937016659 269480 00888937016659 269482 00888937016659 269483 00888937016659 274957 00888937016659 274958 00888937016659 274960 00888937016659 274961 00888937016659 286859 00888937016659 286861 00888937016659 286863 00888937016659 292881 00888937016659 292883 00888937016659 292888 00888937016659 292889 00888937016659 292900 00888937016659 292901 00888937016659 292903 00888937016659 296236 00888937016659 296238 00888937016659 297033 00888937016659 297034 00888937016659 297035 00888937016659 298049 00888937016659 298055 00888937016659 298054 00888937016659 298056 00888937016659 298062 00888937016659 303771 00888937016659 303772 00888937016659 303950 00888937016659 304053 00888937016659 304363 00888937016659 305019 00888937016659 305020 00888937016659 307724 00888937016659 306339 00888937016659 306341 00888937016659 307728 00888937016659 307726 00888937016659 307730 00888937016659 308591 00888937016659 309910 00888937016659 309912 00888937016659 313142 00888937016659 311533 00888937016659 314675 00888937016659 315626 00888937016659 315810 00888937016659 318970 00888937016659 316866 00888937016659 320400 00888937016659

Distribution

Worldwide distribution: US (Nationwide) including Puerto Rico; and countries (foreign) of: Australia, Bahrain, Belgium, Bulgaria, Canada, Colombia, Cyprus, Ecuador, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Malaysia, Malta, Netherlands, New Zealand, Saudi Arabia, Singapore, Turkey, Philippines, Qatar, United Arab Emirates, and United Kingdom.

Quantity

220,500 devices