FDA Enforcement
Class II
Completed
AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode
Recall: Z-0219-2025
·
Reported October 30, 2024
Enforcement
- Recall Number
- Z-0219-2025
- Event ID
- 95414
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- Ad-Tech Medical Instrument Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 30, 2024
- Initiation Date
- September 19, 2024
- Classification Date
- October 24, 2024
- Address
- 400 W Oakview Pkwy, N/A, Oak Creek, WI, 53154-7213, United States
Description
AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode
Reason
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Code Info
UDI/DI 00841823107510, Lot Number 170052 858859860
Distribution
US Nationwide distribution in the states of CA, OH.
Quantity
2 units