FDA Enforcement Class II Ongoing

Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 550 MD, Rx Only

Recall: Z-0218-2026 · Reported October 29, 2025

Enforcement

Recall Number
Z-0218-2026
Event ID
97668
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
UIH Technologies LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 29, 2025
Initiation Date
September 12, 2025
Classification Date
October 21, 2025
Address
9380 Kirby Dr, Houston, TX, 77054-2515, United States

Description

Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 550 MD, Rx Only

Reason

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Code Info

UDI:(01)06971576831036/ Serial Numbers: 353020 353021 353023 353045 353049 354001 354002 354003 354004 354005 354006 354007 354008 354009 354010 354011 354012 354013 354014 354015 354016 354017

Distribution

US: OUS:

Quantity

22 units