FDA Enforcement
Class II
Terminated
Eon 300 system includes the following product reference numbers and names: Reference no. 79345 Glucose (GPSL) Reference no. 79450 Uric acid (AUVD) Reference no. 79435 Triglycerides (TGML) Reference no. 79295 Cholesterol (CHVD) Reference no. 79301 Cholesterol HDL (HDLL) Reference no. 79310 Cholesterol, LDL (LDLL)
Recall: Z-0215-2016
·
Reported November 11, 2015
Enforcement
- Recall Number
- Z-0215-2016
- Event ID
- 72322
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- ELITech Clinical Systems SAS
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 11, 2015
- Initiation Date
- September 22, 2015
- Classification Date
- November 4, 2015
- Termination Date
- March 4, 2016
- Address
- Zone lndustrielle, N/A, SEES, N/A, N/A, France
Description
Eon 300 system includes the following product reference numbers and names: Reference no. 79345 Glucose (GPSL) Reference no. 79450 Uric acid (AUVD) Reference no. 79435 Triglycerides (TGML) Reference no. 79295 Cholesterol (CHVD) Reference no. 79301 Cholesterol HDL (HDLL) Reference no. 79310 Cholesterol, LDL (LDLL)
Reason
ELITech clinical Systems reagents based on Trinder reaction used on the Eon 300 instrument system are recalled due to potential drug interference. No complaints were received.
Code Info
all lots
Distribution
US Nationwide Distribution
Quantity
612 total