FDA Enforcement Class II Terminated

Beckman Coulter Lipase, REF OSR6130 and OSR6230.

Recall: Z-0210-2020 · Reported October 30, 2019

Enforcement

Recall Number
Z-0210-2020
Event ID
83583
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 30, 2019
Initiation Date
August 24, 2018
Classification Date
October 24, 2019
Termination Date
August 2, 2024
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232, United States

Description

Beckman Coulter Lipase, REF OSR6130 and OSR6230.

Reason

NAPQI, a metabolite of Acetaminophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.

Code Info

All lots

Distribution

Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada.

Quantity

20,926 units