Composite Series Skull Clamp, one clamp per tote. The MAYFIELD¿ Skull Clamp (A3059) is a cranial stabilization device, designed to provide rigid skeletal fixation. The Skull Clamp is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary
Enforcement
- Recall Number
- Z-0210-2014
- Event ID
- 66449
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Integra LifeSciences Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 20, 2013
- Initiation Date
- August 29, 2013
- Classification Date
- November 13, 2013
- Termination Date
- January 14, 2014
- Address
- 311 Enterprise Dr, Plainsboro, NJ, 08536-3344, United States
Description
Composite Series Skull Clamp, one clamp per tote. The MAYFIELD¿ Skull Clamp (A3059) is a cranial stabilization device, designed to provide rigid skeletal fixation. The Skull Clamp is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary
An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in a report of patient injury, adverse health consequence, or a delay / prolongation of the surgical procedure.
The following serial numbers are affected by this recall: SS120076, SS120082, SS120083, SS120084, SS120088, SS120089, SS120090, SS120092, SS120093, SS120094, SS120095, SS120111, SS120112, SS120113, SS120114, SS120115, SS120116, SS120117, SS120118, SS120119, SS120120, SS120121, SS120122, SS120123, SS120124, SS120125, SS120126, SS120127, SS120128, SS120130, SS120131, SS120133, SS120134, SS120135, SS120171, SS120252, SS120256, SS120257
US Distribution including the states of AL, AR, CA, CT, FL, GA, IA, IL, MS, NY, OH, SC, TN, and TX.
37 units