FDA Enforcement Class II Terminated

e.cam Gamma Camera, an emission computed tomography system. It is intended to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.

Recall: Z-0208-2016 · Reported November 11, 2015

Enforcement

Recall Number
Z-0208-2016
Event ID
72296
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 11, 2015
Initiation Date
June 26, 2015
Classification Date
November 4, 2015
Termination Date
December 23, 2016
Address
2501 Barrington Rd, N/A, Hoffman Estates, IL, 60192-2061, United States

Description

e.cam Gamma Camera, an emission computed tomography system. It is intended to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.

Reason

Potential failure mode related to the radial brake, where the detector was able to drift, led to a re-design of the brake redundancy circuit. The redesigned Brake Redundancy Circuit will force even a defective brake to engage in the unpowered state by applying a reverse charge to the brake and ensuring the brake has demagnetized.

Code Info

E.cam Model numbers affected : 4380213, 4380221, 5242826, 5077066, 5077074, 5984005, 5989079, 5989087, 5989095, 5991109, 5991117, 5992099, 7324119, 7324135, 7324143, 7324150, 7760809, 7760932, 7760161, 7761161, 7823920, 7823938, 7823946, 7823953, 7823961, 7823979, 10151531, 10151532, 10520745. This recall applies only to ecam systems that were installed before December 31st 2004. Serial Numbers 7438, 7440, 81010, 1011, 8509, 8510, 1, 1007, 1008, 1010, 1019, 1020, 2001, 2004, 2006, 2007, 2009, 2010, 2014, 2017, 2019, 2022, 2023, 2025, 3002, 3006, 3008, 3010, 3011, 3012, 3013, 3014, 3017, 3021, 3022, 3023, 3024, 3029, 3031, 3034, 4001, 4002, 4004, 4008, 4009, 4010, 4011, 4013, 4014, 4016, 1006, 1009, 1015, 1016, 1020, 1021, 1024, 1025, 1029, 1030, 1031, 1032, 1033, 1041, 1044, 1046, 1051, 1056, 1057, 1063, 1068, 1069, 1070, 1071, 10711, 1074, 1076, 1077, 1079, 1084, 1092, 1095, 1096, 1098, 1100, 1102, 1105, 1106, 1111, 1116, 1117, 1118, 1122, 1123, 1124, 1127, 1130, 1132, 1133, 1134, 1135, 1136, 1137, 1142, 1143, 1146, 1148, 1150, 1151, 1153, 1155, 1157, 1158, 1160, 1161, 1162, 1163, 1165, 1166, 1169, 1172, 1174, 1176, 1180, 1181, 1182, 1184, 1186, 1187, 1188, 1189, 1190, 1191, 1192, 1196, 1197, 1199, 1200, 12009, 1201, 1202, 1205, 1206, 1207, 1208, 121010, 1211, 1234, 126110, 21909, 26011, 31609, 31711, 31827, 41207, 62111, 62310, 7030, 71410, 71411, 7440, 7502, 7582, 8149, 82409, 9, 918073, 1002, 1006, 1009, 1013, 1014, 1015, 1016, 1019, 1020, 1021, 1024, 1029, 1034, 1036, 1037, 1038, 1039, 1046, 1049, 1052, 1056, 1058, 1059, 1062, 1063, 1065, 1067, 1071, 1077, 1080, 1081, 1082, 1083, 1088, 1089, 1094, 1095, 1099, 1101, 1103, 1104, 1105, 1107, 1109, 1110, 1113, 1114, 1116, 1118, 1121, 1129, 1130, 1131, 1135, 1136, 1137, 1138, 1139, 1141, 1142, 1143, 2001, 2004, 2006, 2007, 2008, 2010, 2012, 2013, 2014, 2015, 2017, 2019, 2024, 2026, 2027, 2028, 2030, 2036, 2038, 2039, 2041, 2042, 2045, 2047, 2048, 2050, 2051, 2052, 2055, 2058, 2060, 2065, 2066, 2074, 2075, 2076, 2078, 2079, 2080, 2081, 2084, 2090, 2096, 2098, 2099, 2100, 2107, 2114, 2115, 2117, 2118, 2119, 2120, 2122, 2123, 2124, 2125, 2127, 2128, 2130, 3004, 3005, 3007, 3008, 3010, 3015, 3016, 3018, 3020, 3022, 3023, 3026, 3027, 3029, 3032, 3035, 3037, 3039, 3040, 3045, 3047, 3048, 3049, 3050, 3051, 3052, 3055, 3056, 3057, 3058, 3064, 3065, 3067, 3069, 3070, 3071, 3075, 3076, 3077, 3080, 3082, 3083, 3085, 3087, 3088, 3090, 3091, 3092, 3093, 3094, 3096, 3098, 3100, 3101, 3104, 3105, 3106, 3107, 3109, 3112, 3114, 3115, 3116, 3117, 3118, 3120, 3121, 3122, 3123, 3124, 3125, 3132, 3133, 3135, 3137, 3138, 3139, 3140, 3141, 4001, 4002, 4004, 4006, 4007, 4008, 4010, 4011, 4012, 4014, 4015, 4017, 4018, 4019, 4021, 4022, 4023, 4024, 4025, 4026, 4028, 4030, 4034, 4039, 4041, 4044, 4045, 4048, 4049, 4050, 4052, 4053, 4054, 4056, 4057, 4060, 4062, 4063, 4064, 4066, 4067, 4068, 4069, 4070, 4071, 4076, 4077, 4078, 4090, 10209, 10411, 12210, 12310, 31611, 40109, 4082, 4083, 4084, 4086, 4095, 4096, 4097, 4098, 4100, 4107, 4112, 4114, 4116, 4117, 4118, 4119, 4120, 4121, 4122, 4125, 4127, 4129, 4130, 4131, 4132, 4135, 4136, 4137, 4141, 4142, 4143, 4146, 4147, 4148, 4149, 4151, 4152, 4154, 4156, 4158, 4159, 4164, 4166, 4168, 4171, 4173, 4174, 4177, 4178, 4179, 4181, 4182, 4183, 4185, 4186, 4190, 4196, 4197, 4202, 4205, 4206, 4208, 4209, 4214, 4215, 4217, 4218, 4219, 4220, 4221, 4222, 4223, 4225, 4226, 5001, 5002, 5004, 5005, 5006, 5008, 5009, 5011, 5012, 5013, 5016, 5017, 5018, 5019, 5020, 5021, 5022, 5023, 5024, 5025, 5027, 5029, 5031, 5032, 5033, 5036, 5037, 5039, 5040, 5041, 5042, 5043, 5045, 5046, 5048, 5049, 5050, 5051, 5057, 5060, 5063, 5067, 5069, 5070, 5071, 5102, 5107, 5108, 5117, 5123, 6000, 60911, 62211, 83110, 83210, 92110, 4017, 4018, 4019, 4022, 4023, 4025, 4030, 4032, 4033, 4034, 4035, 4037, 4044, 4046, 4049, 5001, 5003, 5004, 5006, 5007, 5008, 5010, 5011, 5012, 5013, 5015, 5016, 5017, 5018, 5020, 5021, 5022, 5024, 5025, 5028, 5031, 5033, 5036, 5037, 5039, 5040, 5043, 3, 6006, 6007, 6009, 6010, 7001, 7004, 7006, 7409, 7501, 7502, 7507, 7509, 6053, 6501, 7386, 7502, 7504, 7505, 7507, 7508, 7510, 7512, 7513, 7515, 7516, 7517, 7518, 7519, 7520, 7521, 7522, 7523, 7525, 7526, 7529, 7531, 7533, 7534, 7535, 7536, 7538, 7539, 7540, 7542, 7545, 7547, 7548, 7549, 7550, 7551, 7553, 7555, 7556, 7557, 7558, 7561, 7562, 7563, 7564, 7566, 7568, 7569, 7573, 7574, 7576, 7577, 7578, 7579, 7581, 7582, 7584, 7587, 7588, 7589, 7590, 7592, 7595, 7596, 7599, 7600, 7602, 7605, 7606, 7608, 7610, 7611, 7614, 7615, 7616, 7617, 7618, 7618*, 7620, 7621, 7622, 7623, 7624, 7626, 7627, 7628, 7629, 7631, 7635, 7636, 7637, 7638, 7651, 7861, 8001, 8003, 8005, 8006, 8007, 8009, 8010, 8011, 8014, 8015, 8017, 8020, 8021, 8023, 8024, 8025, 8027, 8028, 8029, 8030, 8031, 8032, 8035, 8036, 8037, 8039, 8040, 8044, 8045, 8047, 8049, 8050, 8051, 8053, 8055, 8056, 8057, 8058, 8059, 8060, 8063, 8064, 8067, 8069, 8070, 8071, 8072, 8073, 8075, 8078, 8079, 8080, 8081, 8082, 8085, 8087, 8088, 8089, 8090, 8092, 8094, 8095, 8096, 8097, 8098, 8099, 8100, 8102, 8105, 8106, 8108, 8109, 8110, 8111, 8112, 8113, 8114, 8115, 8116, 8117, 8118, 8119, 8120, 8121, 8122, 8124, 8125, 8126, 8127, 8128, 8130, 8131, 8132, 8133, 8135, 8137, 8138, 8141, 8142, 8144, 8145, 8146, 8147, 8501, 8502, 8503, 8504, 8505, 8506, 8508, 8510, 8511, 8512, 8513, 8514, 8515, 8516, 8517, 8518, 8519, 8520, 8521, 8522, 8523, 8524, 8525, 8526, 8527, 8528, 8529, 8530, 8532, 8533, 8534, 8535, 8536, 8538, 8540, 8541, 8543, 8544, 8545, 8548, 8549, 8550, 8552, 8553, 8556, 8558, 8559, 8560, 8561, 8562, 8563, 8564, 8565, 8567, 8568, 8569, 8570, 8571, 8572, 8573, 8574, 8575, 8576, 8577, 8578, 8580, 8581, 8583, 8585, 8586, 8587, 8588, 8589, 8590, 8592, 8593, 8595, 8596, 8598, 8600, 8603, 8604, 8605, 8606, 8607, 8608, 8609, 8610, 8611, 8612, 8613, 8615, 8616, 8617, 8618, 8619, 8620, 8621, 8622, 8624, 8625, 8626, 8628, 8629, 8630, 8631, 8632, 8633, 8634, 8635, 8636, 8637, 8638, 8639, 8640, 8641, 8642, 8643, 8645, 8646, 8647, 8650, 8651, 8652, 8653, 8654, 8655, 8656, 8657, 8658, 8659, 8660, 8661, 8662, 8665, 8666, 8667, 8668, 8669, 8670, 8671, 8672, 8673, 8674, 8675, 8676, 8677, 8678, 8679, 8681, 8683, 8684, 8685, 8686, 8687, 8688, 8689, 8690, 8691, 8692, 8693, 8694, 8695, 8696, 8697, 8698, 8699, 8700, 8701, 8702, 8703, 8704, 8705, 8706, 8707, 8708, 8709, 8710, 8711, 8712, 8714, 8716, 8717, 8718, 8719, 8720, 8721, 8722, 8723, 8724, 8725, 8726, 8727, 8728, 8729, 8730, 8731, 8732, 8733, 8734, 8735, 8736, 8738, 8739, 8740, 8741, 8743, 8744, 8745, 8746, 8747, 8748, 8749, 8750, 8754, 8755, 8756, 8757, 8758, 8759, 8760, 8761, 8763, 8766, 8767, 8769, 8770, 8771, 8773, 8778, 8779, 8780, 8781, 8782, 8783, 8784, 8786, 8788, 8789, 8790, 8791, 8792, 8793, 8794, 8799, 8801, 8802, 8803, 8804, 8805, 8807, 8808, 8809, 8810, 8811, 8812, 8813, 8814, 8816, 8817, 8818, 8819, 8820, 8822, 8823, 8824, 8825, 8827, 8828, 8829, 8830, 8831, 8832, 8833, 8834, 8835, 8836, 8837, 8838, 8839, 8841, 8842, 8844, 5045, 7502, 7503, 7505, 7506, 7509, 7511, 8001, 8003, 8004, 8007, 8009, 8010, 8012, 8013, 8014, 8015, 8016, 8017, 8019, 8021, 8022, 8024, 8025, 8027, 8028, 8030, 8032, 8501, 8503, 8504, 8506, 8507, 8508, 8509, 8511, 8512, 8513, 8514, 8517, 8518, 8519, 8521, 8523, 8524, 8525, 8526, 8528, 8530, 8532, 8534, 8535, 8536, 8537, 8539, 8540, 8543, 8545, 8547, 8548, 8549, 8550, 8552, 8553, 8554, 8555, 8556, 7090, 7502, 7504, 7505, 8001, 8003, 8005, 8021, 8501, 8502, 8503, 8504, 8505, 8506, 1001, 13044, 5039, 5505, 5507, 5508, 5510, 5511, 5512, 5514, 5517, 5518, 5519, 5521, 5522, 5523, 5524, 5525, 5526, 5527, 5528, 5529, 5530, 5531, 5532, 5533, 5535, 5536, 5537, 5538, 5539, 5540, 5541, 5542, 5546, 5547, 5548, 5550, 5551, 5553, 5555, 5556, 5557, 5558, 5559, 5560, 5561, 5562, 5563, 5565, 5567, 5570, 5571, 5572, 5573, 5575, 5578, 5579, 5580, 5581, 5584, 5585, 5586, 5587, 5588, 5589, 5590, 5591, 5593, 5594, 5595, 5597, 5599, 5600, 5601, 5602, 5603, 5604, 5605, 5606, 5607, 5608, 5609, 5610, 5611, 5612, 5613, 5614, 5616, 5617, 5618, 5620, 5622, 5623, 5624, 5625, 5629, 5631, 5632, 5634, 5636, 5637, 5639, 5640, 5642, 5644, 5645, 5646, 5647, 5649, 5650, 5658, 5671, 5672, 5673, 5674, 5675, 5676, 5677, 5678, 5679, 7005, 7006, 7007, 7009, 7011, 7012, 7013, 7014, 7016, 7017, 7018, 7019, 7020, 7022, 7025, 7027, 7028, 7029, 7030, 7031, 7032, 7033, 7034, 7036, 7037, 7039, 7040, 7041, 7043, 7044, 7045, 7046, 7047, 7048, 7049, 7050, 7051, 7056, 7058, 7059, 7060, 7061, 7062, 7063, 7065, 7067, 7068, 7069, 7070, 7073, 7076, 7078, 7080, 7082, 7083, 7084, 7086, 7088, 7090, 7091, 7092, 7094, 7095, 7096, 7097, 7098, 7102, 7103, 7104, 7105, 7109, 7110, 7111, 7112, 7117, 7120, 7121, 7124, 7126, 7127, 7501, 7502, 7503, 7504, 7505, 7506, 7508, 7509, 7510, 7512, 7513, 7515, 7516, 7517, 7519, 7520, 7521, 7523, 7524, 7525, 7526, 7530, 7531, 7532, 7533, 7534, 7535, 7536, 7537, 7538, 7539, 7541, 7544, 7545, 7551, 7552, 7553, 7555, 7556, 7557, 7560, 7561, 7562, CAX271/U2B004, 5501, 5502, 5503, 5506, 5511, 5512, 5513, 5514, 5516, 5517, 5519, 5521, 5523, 5525, 5526, 5527, 5529, 7003, 7004, 7005, 7006, 7008, 7011, 7013, 7015, 7017, 7018, 7019, 7020, 7021, 7022, 7024, 7025, 7026, 7501, 7504, 7506, 7507, 7508, 7509, 7513, 5101, 5102, 5104, 5109, 5110, 5112, 5113, 5115, 5116, 5118, 5121, 5125, 5126, 5127, 5128, 5131, 5132, 5133, 5135, 5136, 5142, 5144, 5146, 5147, 5148, 5149, 5150, 5151, 5153, 5157, 5158, 5159, 5160, 5162, 5165, 5167, 5168, 5170, 5171, 5172, 5173, 5174, 5502, 1003, 548, 550, 558, 564, 579, 581, 588, 589, 590, 592, 593, 595, RC528, RC537, RC549, 551, 584, RC521, RC536, 10610, 21511, 510, 529, 530, 557, 559, 573, 62310, RC498, RC543, 556, 572, RC493, RC524, RC538, RC541, RC545, RC546, 8501, 8503, 8504, 8505, 8506, 8508, 8509, 8510, 8511, 100003, 1234, 1234578, 4916, 60151, 63010, 7409, 7502, 8191, 8501, 8502, 8503, 8506, 8507, 8508, 8509, 8512, 8514, 8515, 8517, 8519, 8520, 8521, 8522, 8523, 8524, 8525, 8529, 8530, 8531, 8532, 8533, 8534, 8537, 8538, 8539, 8540, 8541, 8542, 8543, 8544, 8546, 8547, 8548, 8549, 8550, 8551, 8552, 8553, 8554, 8555, 8556, 8557, 8560, 8561, 8562, 8563, 8564, 8565, 8566, 8567, 8568, 8569, 8570, 8571, 8572, 8573, 8574, 8577, 8578, 8579, 8580, 8581, 8583, 9003, 9004, 9006, 9007, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9022, 9023, 9024, 9026, 9027, 9028, 9029, 9030, 9031, 9032, 9033, 9034, 9035, 9036, 9037, 9038, 9039, 9041, 9043, 8473046345, 8503, 8205, 8505, 8506, 8509, 8510, 8511, 8513, 8514, 8515, 8516, 8517, 8518, 8519, 8520, 9821, 8501, 8501*, 8502, 8507, 8508, 10000, 10001, 10002, 10003, 10005, 10006, 10007, 10008, 10009, 10010, 10013, 10014, 10015, 10016, 10017, 10019, 10020, 10022, 10023, 10024, 10025, 10026, 10028, 10029, 10030, 10031, 10032, 10033, 10034, 10035, 10036, 10037, 10038, 10040, 10041, 10042, 10044, 10046, 10047, 10048, 10049, 10050, 10051, 10052, 10053, 10054, 10055, 10056, 10057, 10058, 10059, 10060, 10061, 10062, 10063, 10065, 10066, 10067, 10068, 10069, 10070, 10071, 10072, 10073, 10074, 10076, 10077, 10078, 10079, 10080, 10082, 10083, 10084, 10085, 10087, 10088, 10089, 10090, 10091, 10092, 10093, 10094, 10096, 10098, 10099, 10100, 10101, 10102, 10103, 10104, 10106, 10107, 10108, 10109, 10111, 10112, 10113, 10114, 10115, 10117, 10118, 10119, 10120, 10121, 10122, 10124, 10125, 10126, 10127, 10128, 10129, 10130, 10131, 10132, 10133, 10135, 10136, 10137, 10138, 10139, 10140, 10141, 10142, 10143, 10144, 10147, 10149, 10150, 10151, 10153, 10154, 10155, 10156, 10157, 10158, 10159, 10160, 10161, 10162, 10163, 10164, 10165, 10166, 10167, 10168, 10169, 10170, 10171, 10172, 10173, 10174, 10175, 10176, 10177, 10178, 10179, 10180, 10181, 10182, 10183, 10184, 10185, 10186, 10187, 10188, 10191, 10192, 10193, 10194, 10195, 10198, 10199, 10200, 10201, 10202, 10203, 10204, 10205, 10206, 10207, 10208, 10209, 10211, 10212, 10213, 10214, 10215, 10216, 10217, 10218, 10219, 10220, 10221, 10222, 10224, 10225, 10226, 10227, 10228, 10230, 10231, 10232, 10233, 10234, 10235, 10236, 10237, 10238, 10239, 10240, 10241, 10243, 10245, 10246, 10248, 10249, 10250, 10251, 10252, 10253, 10254, 10255, 10256, 10257, 10258, 10259, 10260, 10261, 10262, 10263, 10264, 10265, 10266, 10267, 10268, 10269, 10270, 10271, 10272, 10273, 10274, 10275, 10276, 10277, 10278, 10280, 10282, 10283, 10284, 10285, 10286, 10287, 10288, 10289, 10290, 10291, 10292, 10293, 10294, 10295, 10296, 10297, 10298, 10299, 10300, 10301, 10302, 10303, 10304, 10305, 10306, 10307, 10308, 10309, 10310, 10311, 10312, 10313, 10314, 10315, 10317, 10318, 10319, 10320, 10321, 10322, 10323, 10324, 10325, 10326, 10327, 10328, 10329, 10330, 10331, 10332, 10333, 10334, 10336, 10337, 10338, 10339, 10340, 10341, 10343, 10344, 5382, 5517, 5764, 8490, 9001, 9002, 9003, 9004, 9005, 9006, 9008, 9009, 9010, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9022, 9023, 9024, 9025, 9026, 9027, 9028, 9031, 9033, 9034, 9035, 9036, 9038, 9039, 9040, 9041, 9043, 9044, 9045, 9046, 9047, 9049, 9050, 9051, 9053, 9054, 9055, 9057, 9058, 9059, 9060, 9061, 9062, 9063, 9064, 9066, 9068, 9069, 9070, 9073, 9074, 9076, 9077, 9078, 9079, 9080, 9081, 9082, 9083, 9084, 9085, 9086, 9087, 9089, 9090, 9091, 9092, 9094, 9095, 9096, 9097, 9098, 9099, 9100, 9101, 9102, 9103, 9104, 9106, 9107, 9108, 9109, 9110, 9111, 9112, 9113, 9114, 9115, 9116, 9117, 9120, 9122, 9123, 9124, 9125, 9126, 9127, 9129, 9130, 9132, 9133, 9134, 9135, 9136, 9137, 9138, 9139, 9142, 9143, 9144, 9145, 9146, 9147, 9148, 9149, 9150, 9151, 9152, 9154, 9155, 9156, 9157, 9158, 9159, 9161, 9162, 9163, 9164, 9165, 9166, 9167, 9169, 9170, 9171, 9172, 9173, 9174, 9175, 9176, 9177, 9179, 9180, 9181, 9182, 9183, 9184, 9185, 9186, 9187, 9189, 9190, 9191, 9193, 9194, 9198, 9199, 9200, 9201, 9204, 9206, 9207, 9210, 9212, 9213, 9214, 9215, 9216, 9219, 9220, 9221, 9223, 9224, 9225, 9226, 9227, 9228, 9230, 9231, 9232, 9233, 9234, 9235, 9238, 9239, 9240, 9241, 9242, 9243, 9245, 9246, 9247, 9248, 9251, 9252, 9253, 9254, 9255, 9256, 9259, 9260, 9261, 9262, 9263, 9264, 9265, 9266, 9267, 9268, 9269, 9270, 9271, 9272, 9273, 9274, 9275, 9276, 9277, 9279, 9280, 9281, 9282, 9283, 9284, 9285, 9287, 9288, 9289, 9290, 9292, 9293, 9294, 9295, 9296, 9297, 9298, 9299, 9300, 9301, 9302, 9303, 9304, 9305, 9306, 9307, 9308, 9309, 9311, 9313, 10001, 10002, 10003, 10004, 10005, 10006, 10007, 10008, 10009, 10011, 10012, 10014, 10015, 10016, 10017, 10018, 10019, 10022, 10023, 10025, 10026, 10027, 10028, 10029, 10030, 10031, 10032, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10044, 10045, 10046, 10047, 10048, 10049, 10050, 10051, 10052, 10055, 10056, 10057, 10058, 10059, 10061, 8505, 8509, 9001, 9002, 9003, 9004, 9005, 9007, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9019, 9020, 9021, 9022, 9024, 9026, 9031, 9032, 9033, 9034, 9036, 9037, 9038, 9039, 9040, 9041, 9042, 9043, 9044, 9045, 9001, 9002, 9005, 9006, 9007, 9009, 9010, 9011, 9013, 9014, 9016, and 9017.

Distribution

Worldwide Distribution: US (nationwide) and countries of: United Arab Emirates, Albania, Argentina, Austria, Australia, Bosnia and Herzegovina, Bangladesh, Belgium, Bolivia, Brazil, Belarus, Canada, Switzerland, Chile, Cameroon, China, Colombia, Costa Rica, Czech Republic, Germany, Denmark, Dominica, Ecuador, Egypt, Spain, Finland, France, United Kingdom, Georgia, Greece, Croatia, Canary Islands, Ireland, Israel, India, Iraq, Iran, Islamic-Republic of, Italy, Jordan, Japan, Korea-Republic of, Kuwait, Lebanon, Sri Lanka, Luxembourg, Latvia, Morocco, Myanmar, Mauritius, Mexico, Malaysia, Netherlands, Norway, Nepal, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Portugal, Qatar, Serbia, Russian Federation, Saudi Arabia, Sweden, Slovakia, El Salvador, Syrian Arab Republic (Syria), Thailand, Tunisia, Turkey, Taiwan, Republic of China, Ukraine, Uganda, Uzbekistan, and South Africa.

Quantity

32296 e.cam Gamma Camera systems