FDA Enforcement Class II Ongoing

medex TranStar MONITORING SET W/SAMPLING PORT 10/EA, Product Code REF MXG300260

Recall: Z-0204-2025 · Reported October 30, 2024

Enforcement

Recall Number
Z-0204-2025
Event ID
95438
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smiths Medical ASD, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
October 30, 2024
Initiation Date
September 24, 2024
Classification Date
October 23, 2024
Address
6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States

Description

medex TranStar MONITORING SET W/SAMPLING PORT 10/EA, Product Code REF MXG300260

Reason

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Code Info

UDI/DI 50351688518833 (case), 10351688518835 (pouch), Lot Numbers: 4270120, 4297963, 4316640, 4319604, 4323118, 4287727, 4359814, 4387976, 4340473

Distribution

Worldwide distribution.

Quantity