FDA Enforcement
Class I
Terminated
ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code PLEE60A
Recall: Z-0202-2020
·
Reported November 6, 2019
Enforcement
- Recall Number
- Z-0202-2020
- Event ID
- 83897
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ethicon Endo-Surgery Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 6, 2019
- Initiation Date
- October 3, 2019
- Classification Date
- October 29, 2019
- Termination Date
- June 3, 2021
- Address
- 4545 Creek Rd, Blue Ash, OH, 45242-2803, United States
Description
ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code PLEE60A
Reason
The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.
Code Info
Lot: T93Z1G, T93Y4M, T94045, T94117, T93X17, T94087, T93Z2X, T9405L, T94253, T93X95, T93Z75, T93Z2W, T9413Z
Distribution
Worldwide distribution. US Nationwide. Belgium, Brazil, Canada, Chile, Colombia, Japan, Kuwait, Saudi Arabia, and Singapore.
Quantity
5040 units