FDA Enforcement Class I Terminated

ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60A

Recall: Z-0201-2020 · Reported November 6, 2019

Enforcement

Recall Number
Z-0201-2020
Event ID
83897
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Ethicon Endo-Surgery Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 6, 2019
Initiation Date
October 3, 2019
Classification Date
October 29, 2019
Termination Date
June 3, 2021
Address
4545 Creek Rd, Blue Ash, OH, 45242-2803, United States

Description

ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60A

Reason

The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

Code Info

Lot: T93Z9Y, T9411A

Distribution

Worldwide distribution. US Nationwide. Belgium, Brazil, Canada, Chile, Colombia, Japan, Kuwait, Saudi Arabia, and Singapore.

Quantity

255 units