FDA Enforcement Class II Ongoing

Beaver Arthro-Lok Pointed Tip Curve L 4mm Blade-an orthopedic blade, which may be used in various types of orthopedic surgery (i.e. foot, ankle, hand, knee, hip or spine surgery). Catalog#: 376780

Recall: Z-0197-2023 · Reported November 16, 2022

Enforcement

Recall Number
Z-0197-2023
Event ID
91005
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beaver Visitec International, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
November 16, 2022
Initiation Date
September 30, 2022
Classification Date
November 7, 2022
Address
500 Totten Pond Rd, 10 City Point, Waltham, MA, 02451-1916, United States

Description

Beaver Arthro-Lok Pointed Tip Curve L 4mm Blade-an orthopedic blade, which may be used in various types of orthopedic surgery (i.e. foot, ankle, hand, knee, hip or spine surgery). Catalog#: 376780

Reason

Incorrect configuration of the blades in the package. The blades curve to the right instead of curving to the left. If used may cause cause unintended tissue damage during the procedure or if noticed prior to use may cause a delay in surgery

Code Info

UDI# 00886158001249 LOT#'s: 6034831,6039837

Distribution

FL, GA, IN, LA, MA, ME, VA Foreign: France, Japan, Netherlands, Switzerland, United Kingdom

Quantity

150 units US; 1116 units OUS