7" (18 cm) Appx 0.31 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave Clear, Clamp, NanoClave T-Connector (Purple Ring), Rotating Luer; unit per pouch, 50 multivac pouches per case. UDI:(01)10887709088190(17)241201(30)50(10)4515232;(01)10887709068350(17)250101(30)50(10)4572761 The high pressure extension set is a device used as conduit tubing to deliver fluids from an ~f electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices.
Enforcement
- Recall Number
- Z-0197-2021
- Event ID
- 86054
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- ICU Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 28, 2020
- Initiation Date
- July 6, 2020
- Classification Date
- October 22, 2020
- Termination Date
- May 11, 2023
- Address
- 951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States
Description
7" (18 cm) Appx 0.31 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave Clear, Clamp, NanoClave T-Connector (Purple Ring), Rotating Luer; unit per pouch, 50 multivac pouches per case. UDI:(01)10887709088190(17)241201(30)50(10)4515232;(01)10887709068350(17)250101(30)50(10)4572761 The high pressure extension set is a device used as conduit tubing to deliver fluids from an ~f electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices.
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Lot Numbers:4515232, 4572761.
Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
N/A