FDA Enforcement Class II Terminated

The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.

Recall: Z-0196-2015 · Reported November 19, 2014

Enforcement

Recall Number
Z-0196-2015
Event ID
69472
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Steris Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 19, 2014
Initiation Date
October 10, 2014
Classification Date
November 7, 2014
Termination Date
June 26, 2015
Address
5960 Heisley Rd, Mentor, OH, 44060-1834, United States

Description

The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.

Reason

The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.

Code Info

5085 and 5085SRT

Distribution

US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.

Quantity

999