FDA Enforcement
Class II
Terminated
The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.
Recall: Z-0196-2015
·
Reported November 19, 2014
Enforcement
- Recall Number
- Z-0196-2015
- Event ID
- 69472
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Steris Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 19, 2014
- Initiation Date
- October 10, 2014
- Classification Date
- November 7, 2014
- Termination Date
- June 26, 2015
- Address
- 5960 Heisley Rd, Mentor, OH, 44060-1834, United States
Description
The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.
Reason
The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.
Code Info
5085 and 5085SRT
Distribution
US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.
Quantity
999