FDA Enforcement Class II Ongoing

medex TranStar DPT w/Stopcock & 3cc Flush 10/EA, Product Code REF MX950SCFT

Recall: Z-0193-2025 · Reported October 30, 2024

Enforcement

Recall Number
Z-0193-2025
Event ID
95438
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smiths Medical ASD, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
October 30, 2024
Initiation Date
September 24, 2024
Classification Date
October 23, 2024
Address
6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States

Description

medex TranStar DPT w/Stopcock & 3cc Flush 10/EA, Product Code REF MX950SCFT

Reason

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Code Info

UDI/DI 50351688503945 (case), 10351688503947 (pouch), Lot Numbers: 4277663

Distribution

Worldwide distribution.

Quantity

10 units