FDA Enforcement
Class II
Ongoing
Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1 and lower. SMN 10703038
Recall: Z-0193-2022
·
Reported November 10, 2021
Enforcement
- Recall Number
- Z-0193-2022
- Event ID
- 88865
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 10, 2021
- Initiation Date
- October 6, 2021
- Classification Date
- November 2, 2021
- Address
- 500 Gbc Dr, N/A, Newark, DE, 19702-2466, United States
Description
Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1 and lower. SMN 10703038
Reason
The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical interpretations and clinical decisions.
Code Info
UDI 00630414596754
Distribution
Worldwide distribution.
Quantity
1142