pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blisters per box), STERILE R, Rx ONLY The DORO(R) Skull Pins are inserted into the receptacles of the skull clamp and fixate the head during surgery.
Recall: Z-0193-2018
·
Reported December 13, 2017
Enforcement
- Recall Number
- Z-0193-2018
- Event ID
- 78501
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Pro-Med Instruments Gmbh
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 13, 2017
- Initiation Date
- November 2, 2017
- Classification Date
- December 5, 2017
- Termination Date
- July 11, 2018
- Address
- Botzinger Str. 38, N/A, Freiburg Im Breisgau, N/A, N/A, Germany
Description
pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blisters per box), STERILE R, Rx ONLY The DORO(R) Skull Pins are inserted into the receptacles of the skull clamp and fixate the head during surgery.
Reason
The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at the very distal end of the conical part of the skull pin tip.
Code Info
Lots: 17041 and 17272
Distribution
Worldwide Distribution - US (nationwide) Internationally to Germany, Saudi Arabia, France, United Kingdom, Australia, Austria, Ireland, South Africa, Hungary, India, Spain, Italy, Japan, and Canada
Quantity
1142 units