Monarch Airway Clearance System, Model 1000, Product Numbers: PMACS1NA (has blue tooth capability only) and PMACS1WI (has WIFI capability). The firm name on the label is Hill-Rom Services Pte. Ltd., Singapore. The Monarch Airway Clearance System, Model 1000, is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physicians choice of treatment. It is indicated for patients having difficulty with secretion clearance or the presence of atelectasis caused by mucus plugging.
Enforcement
- Recall Number
- Z-0192-2018
- Event ID
- 78100
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- HILL-ROM MANUFACTURING, INC.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- December 6, 2017
- Initiation Date
- September 15, 2017
- Classification Date
- November 30, 2017
- Termination Date
- May 29, 2019
- Address
- 1020 County Road F W, N/A, Saint Paul, MN, 55126-2910, United States
Description
Monarch Airway Clearance System, Model 1000, Product Numbers: PMACS1NA (has blue tooth capability only) and PMACS1WI (has WIFI capability). The firm name on the label is Hill-Rom Services Pte. Ltd., Singapore. The Monarch Airway Clearance System, Model 1000, is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physicians choice of treatment. It is indicated for patients having difficulty with secretion clearance or the presence of atelectasis caused by mucus plugging.
Monarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.
Serial numbers S082VM0092, S122VM0171, S128VM0205, S151VM0242, S082VM0093, S122VM0172, S132VM0210, S151VM0243, S082VM0095, S122VM0175, S132VM0212, S152VM0251, S095VM0119, S122VM0176, S132VM0213, S157VM0257, S095VM0124, S124VM0177, S135VM0216, S157VM0261, S095VM0125, S124VM0178, S136VM0219, S158VM0264, S114VM0152, S124VM0180, S136VM0222, S158VM0266, S119VM0158, S124VM0181, S136VM0223, S158VM0267, S119VM0160, S126VM0187, S138VM0224, S158VM0268, S119VM0161, S126VM0193, S138VM0226, S158VM0271, S119VM0162, S126VM0198, S146VM0227, S158VM0273, S122VM0165, S126VM0199, S146VM0229, S159VM0276, S122VM0166, S128VM0202, S146VM0230, S160VM0279, S122VM0168, S128VM0203, S151VM0238, S167VM0289, S122VM0170, S128VM0204, S151VM0240, and S167VM0291.
USA (nationwide) Distribution to the states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, MA, MI, MN, MO, NE, OH, OK, OR, TN, TX, VA, WA, WI, and WV. Internationally to Canada. There was no military/government distribution.
60 devices