Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3
Enforcement
- Recall Number
- Z-0190-2022
- Event ID
- 88840
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- FujiFilm Healthcare Americas Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 10, 2021
- Initiation Date
- October 14, 2021
- Classification Date
- November 2, 2021
- Address
- 1959 Summit Commerce Park, N/A, Twinsburg, OH, 44087-2371, United States
Description
Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
Serial Numbers: 205W3386 (01)04562122958974(11)171114(21)205W3386; 206F7900 (01)04562122958974(11)180926(21)206F7900; 206F7901 (01)04562122958974(11)180926(21)206F7901; 206F7936 (01)04562122958974(11)181113(21)206F7936; 206F7937 (01)04562122958974(11)181113(21)206F7937; 206F7940 (01)04562122958974(11)181113(21)206F7940; 206F7941 (01)04562122958974(11)181113(21)206F7941; 206F7942 (01)04562122958974(11)181113(21)206F7942; 206F7944 (01)04562122958974(11)181113(21)206F7944; 206F7945 (01)04562122958974(11)181113(21)206F7945; 206F7946 (01)04562122958974(11)181113(21)206F7946; 206F7947 (01)04562122958974(11)181114(21)206F7947; G3002552 (01)04562122958974(11)20190115(21)G3002552; G3024545 (01)04562122958974(11)190607(21)G3024545; G3037885 (01)04562122958974(11)20190719(21)G3037885; G3037887 (01)04562122958974(11)20190719(21)G3037887; G3042405 (01)04562122958974(11)190829(21)G3042405; G3042823 (01)04562122958974(11)20190807(21)G3042823; G3042828 (01)04562122958974(11)20190808(21)G3042828; G3045094 (01)04562122958974(11)20190822(21)G3045094; G3052105 (01)04562122958974(11)20191008(21)G3052105 G3052118 (01)04562122958974(11)20191025(21)G3052118 G3055549 (01)04562122958974(11)20191118(21)G3055549 G3055550 (01)04562122958974(11)20191118(21)G3055550 G3055552 (01)04562122958974(11)20191118(21)G3055552 G3055553 (01)04562122958974(11)20191118(21)G3055553 G3055554 (01)04562122958974(11)20191118(21)G3055554 G3055555 (01)04562122958974(11)20191118(21)G3055555 G3060234 (01)04562122958974(11)20191206(21)G3060234 G3060254 (01)04562122958974(11)20191206(21)G3060254 G3060275 (01)04562122958974(11)20191216(21)G3060275 G3060278 (01)04562122958974(11)20191216(21)G3060278 G3060690 (01)04562122958974(11)200127(21)G3060690 G3071074 (01)04562122958974(11)20200227(21)G3071074 G3080785 (01)04562122958974(11)201105(21)G3080785 G3090450 (01)04562122958974(11)201225(21)G3090450 G3110884 (01)04562122958974(11)210401(21)G3110884
Nationwide
38 units