FDA Enforcement Class II Terminated

SIEMENS N Protein Standard SL, for use on the BN II, BN ProSpec Systems, REF/Product Code OQIM 13, Siemens Material Number (SMN)10446073

Recall: Z-0190-2018 · Reported December 6, 2017

Enforcement

Recall Number
Z-0190-2018
Event ID
77515
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
CSL Behring GmbH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 6, 2017
Initiation Date
June 5, 2017
Classification Date
November 30, 2017
Termination Date
April 7, 2020
Address
Emil-von-Behring-Str. 76, N/A, Marburg, N/A, N/A, Germany

Description

SIEMENS N Protein Standard SL, for use on the BN II, BN ProSpec Systems, REF/Product Code OQIM 13, Siemens Material Number (SMN)10446073

Reason

Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

Code Info

Lot Numbers(UDI): 083611A (00842768007590083611A20170711), 083611B (00842768007590083611B20170711), 083611C (00842768007590083611C20170711), 083611D (00842768007590083611D20170711), exp. date 2017-07-11; 083612 (0084276800759008361220171014), 083612A (00842768007590083612A20171014), 083612B (00842768007590083612B20171014), exp. date 2017-10-14; 083614A (00842768007590083614A20180127), 083614B (00842768007590083614B20180127), 083614C (00842768007590083614C20180127), exp. date 2018-01-27; 083615 (0084276800759008361520180411), 083615A (00842768007590083615A20180411), 083615B (00842768007590083615B20180411), 083615C (00842768007590083615C20180411), exp. date 2018-04-11; 083616B (00842768007590083616B20180730), 083616C (00842768007590083616C20180730), 083616D (00842768007590083616D20180730), exp. date 2018-07-30.

Distribution

Worldwide including the countries of United States, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Great Britain, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, japan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Republic of Korea (S), Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, South Africa, Switzerland, Syria, Taiwan, Thailand, Trinidad-Tobago, Turkey, Turkmenistan, United Arab Emirates, Uruguay, Vatican, Venezuela,

Quantity

N/A