Torrent Suite Dx Software Version 5.8 Catalog Number:A36601 and A36602 OEM Customer software: Torrent Suite Dx T430 Software v5.0 DARUI DR-SEQ 800 Software UDI:(01)10190302014090 (11)000000(10)5.8 (240)A36601
Enforcement
- Recall Number
- Z-0188-2020
- Event ID
- 83786
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Life Technologies Holdings Pte Ltd
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 30, 2019
- Initiation Date
- September 26, 2019
- Classification Date
- October 23, 2019
- Termination Date
- November 23, 2021
- Address
- Blk 33 Marsiling Ind Estate Road 3, #07-06, Singapore, N/A, Singapore
Description
Torrent Suite Dx Software Version 5.8 Catalog Number:A36601 and A36602 OEM Customer software: Torrent Suite Dx T430 Software v5.0 DARUI DR-SEQ 800 Software UDI:(01)10190302014090 (11)000000(10)5.8 (240)A36601
Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant
The field action is associated with the software available on the PGM Dx instrument, therefore no lot code is applicable for this part type
states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland
total of Ion+vela +chassis =400 units