FDA Enforcement
Class II
Ongoing
Medical convenience kits IHDC30R D AND C PACK
Recall: Z-0187-2026
·
Reported October 22, 2025
Enforcement
- Recall Number
- Z-0187-2026
- Event ID
- 97659
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- American Contract Systems Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 22, 2025
- Initiation Date
- September 4, 2025
- Classification Date
- October 16, 2025
- Address
- 1601 Se Gateway Dr Ste 120, Grimes, IA, 50111-5223, United States
Description
Medical convenience kits IHDC30R D AND C PACK
Reason
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Code Info
IHDC30R UDI-DI 191072226310 lot 8096511 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Quantity
32,433 total