FDA Enforcement Class I Terminated

LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.

Recall: Z-0187-2018 · Reported December 20, 2017

Enforcement

Recall Number
Z-0187-2018
Event ID
78303
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Physio-Control, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Press Release
Report Date
December 20, 2017
Initiation Date
October 27, 2017
Classification Date
December 12, 2017
Termination Date
June 12, 2019
Address
11811 Willows Rd NE, N/A, Redmond, WA, 98052-2003, United States

Description

LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.

Reason

The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.

Code Info

45932237, 46042286, 45979590, 46050960, 45979954, 46052545, 46007867, 46061770, 46023185, 46063054, 46023823, 46078012,

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico and Guam and countries of: Canada, Austria, Belgium, Czech Republic; Denmark, Finland, France, French Polynesia, Germany, Hungary, Ireland, Italy, Latvia, Netherlands, Australia, China, Hong Kong, Japan, Argentina, Mexico, Norway, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and New Zealand.

Quantity

14,219 units in total