FDA Enforcement
Class II
Ongoing
medex TranStar Kit w/ Stopcock1/EA, Product Code REF MX20740
Recall: Z-0186-2025
·
Reported October 30, 2024
Enforcement
- Recall Number
- Z-0186-2025
- Event ID
- 95438
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Smiths Medical ASD, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- N/A
- Report Date
- October 30, 2024
- Initiation Date
- September 24, 2024
- Classification Date
- October 23, 2024
- Address
- 6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States
Description
medex TranStar Kit w/ Stopcock1/EA, Product Code REF MX20740
Reason
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Code Info
UDI/DI 50351688508957 (case), 10351688508959 (pouch), Lot Numbers: 4273609
Distribution
Worldwide distribution.
Quantity
260 units