FDA Enforcement Class II Terminated

Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator.

Recall: Z-0186-2022 · Reported November 10, 2021

Enforcement

Recall Number
Z-0186-2022
Event ID
88787
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cook Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 10, 2021
Initiation Date
September 28, 2021
Classification Date
October 29, 2021
Termination Date
February 28, 2024
Address
750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States

Description

Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator.

Reason

Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.

Code Info

lot 13846597

Distribution

Domestic distribution to AR AZ CT FL GA IL LA MA ME MN ND NJ NY OH PA TX WA. Foreign distribution to Germany, Finland, Italy, Netherlands, and Switzerland.

Quantity

48 (US)