FDA Enforcement Class II Terminated

Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.6.15

Recall: Z-0186-2020 · Reported October 30, 2019

Enforcement

Recall Number
Z-0186-2020
Event ID
83786
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Life Technologies Holdings Pte Ltd
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 30, 2019
Initiation Date
September 26, 2019
Classification Date
October 23, 2019
Termination Date
November 23, 2021
Address
Blk 33 Marsiling Ind Estate Road 3, #07-06, Singapore, N/A, Singapore

Description

Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.6.15

Reason

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

Code Info

The field action is associated with the software available on the PGM Dx instrument, therefore no lot code is applicable for this part type

Distribution

states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland

Quantity

total of Ion+vela +chassis =400 units