FDA Enforcement Class II Terminated

Combiset Blood Line with BVM for Hemodialysis use Catalog Number: 03-2965-9. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.

Recall: Z-0185-2014 · Reported November 20, 2013

Enforcement

Recall Number
Z-0185-2014
Event ID
65739
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Fresenius Medical Care Holdings, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 20, 2013
Initiation Date
August 29, 2013
Classification Date
November 8, 2013
Termination Date
February 11, 2021
Address
920 Winter St, Waltham, MA, 02451-1521, United States

Description

Combiset Blood Line with BVM for Hemodialysis use Catalog Number: 03-2965-9. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.

Reason

Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss

Code Info

All lot codes

Distribution

Worldwide Distribution-USA (nationwide) and the country of Canada.

Quantity

24,202 cases