FDA Enforcement Class II Ongoing

Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACK

Recall: Z-0182-2026 · Reported October 22, 2025

Enforcement

Recall Number
Z-0182-2026
Event ID
97659
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
American Contract Systems Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 22, 2025
Initiation Date
September 4, 2025
Classification Date
October 16, 2025
Address
1601 Se Gateway Dr Ste 120, Grimes, IA, 50111-5223, United States

Description

Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACK

Reason

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Code Info

LKRA40G UDI-DI 191072189691 Lot 8025211 Bag serial number 68154347 * If bag serial number is listed as "N/A", then whole lot is affected

Distribution

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Quantity

32,433 total