FDA Enforcement Class II Terminated

Philips Medical Systems Allura Xper FD20C Radiological Imaging with software version R8.2.O; System Code: 722028. The Allura Xper FD series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.

Recall: Z-0181-2015 · Reported November 19, 2014

Enforcement

Recall Number
Z-0181-2015
Event ID
68723
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 19, 2014
Initiation Date
July 8, 2014
Classification Date
November 7, 2014
Termination Date
January 18, 2017
Address
3000 Minuteman Rd, Andover, MA, 01810-1032, United States

Description

Philips Medical Systems Allura Xper FD20C Radiological Imaging with software version R8.2.O; System Code: 722028. The Allura Xper FD series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.

Reason

Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.

Code Info

Allura Xper Software version: R8.2.0

Distribution

Distributed in the states of IL, and TN, and the countries of Argentina , Australia, Belgium, Denmark, Egypt, France, Germany, Hong Kong, Italy, Japan, Korea, Republic of Netherlands, Norway, Saudi Arabia, Switzerland, United Kingdom, and Viet Nam.

Quantity

22 systems