FDA Enforcement Class II Terminated

Reverse Shoulder Prosthesis Stem Impaction Fixture REF 804-03-053 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.

Recall: Z-0180-2016 · Reported November 4, 2015

Enforcement

Recall Number
Z-0180-2016
Event ID
72229
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Encore Medical, Lp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 4, 2015
Initiation Date
September 14, 2015
Classification Date
October 28, 2015
Termination Date
November 4, 2016
Address
9800 Metric Blvd, N/A, Austin, TX, 78758-5445, United States

Description

Reverse Shoulder Prosthesis Stem Impaction Fixture REF 804-03-053 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.

Reason

During the assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.

Code Info

Lot # 52748L02A

Distribution

Nationwide Distribution

Quantity

634 units