FDA Enforcement Class II Ongoing

Medical convenience kits Item Number/Description RCEN25B ENT SDSC 0246972 RCHN27D HEAD AND NECK RCH 0246951

Recall: Z-0178-2026 · Reported October 22, 2025

Enforcement

Recall Number
Z-0178-2026
Event ID
97659
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
American Contract Systems Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 22, 2025
Initiation Date
September 4, 2025
Classification Date
October 16, 2025
Address
1601 Se Gateway Dr Ste 120, Grimes, IA, 50111-5223, United States

Description

Medical convenience kits Item Number/Description RCEN25B ENT SDSC 0246972 RCHN27D HEAD AND NECK RCH 0246951

Reason

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Code Info

RCEN25B UDI-DI 191072225115 lot 8138311 bag serial number N/A; RCHN27D UDI-DI 191072235923 lot 8508211 bag serial number 47847459 * If bag serial number is listed as "n/a", then entire lot is affected

Distribution

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Quantity

32,433 total