Calcium Arsenazo, CALA-0250, For the Quantitative Determination of Total Calcium in Human Serum. The kit label shows CALCIUM ARSENAZO, CALA-0250, R: 12x20 mL. The vial's label shows CALCIUM ARSENAZO, CALA-5220, 20 mL, . CALCIUM ARSENAZO is intended use for the quantitative in vitro diagnostic determination of total calcium in human serum and plasma on ELITech Clinical Systems Selectra Pro Series Analyzers. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). It is not intended for use in Point of Care setting.
Enforcement
- Recall Number
- Z-0178-2014
- Event ID
- 66331
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- SEPPIM SAS
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 13, 2013
- Initiation Date
- September 4, 2013
- Classification Date
- November 7, 2013
- Termination Date
- December 12, 2013
- Address
- ZONE INDUSTRIELLE, N/A, SEES, N/A, N/A, France
Description
Calcium Arsenazo, CALA-0250, For the Quantitative Determination of Total Calcium in Human Serum. The kit label shows CALCIUM ARSENAZO, CALA-0250, R: 12x20 mL. The vial's label shows CALCIUM ARSENAZO, CALA-5220, 20 mL, . CALCIUM ARSENAZO is intended use for the quantitative in vitro diagnostic determination of total calcium in human serum and plasma on ELITech Clinical Systems Selectra Pro Series Analyzers. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). It is not intended for use in Point of Care setting.
Calcium Arsenazo Reagent, reference CALA-0250, is recalled due to complaint of plasma specimens report high results than serum.
Lot numbers distributed in the US: Lot 11-2614, Expiration 07/2013. Lot 12-1002; Expiration 03/2014. Lot 12-2220, Expiration 08/2014. Lot numbers distributed in Sweden: 11-2129, 12-2616, 11-2769, 12-2975, 10-3036, 10-0454, 10-1874, 11-0176, 11-1248, 11-1690, 12-0140, 12-0938, 12-1383, 13-0004, and 13-0408
Worldwide Distribution - US including the states of KY and UT and the country of Sweden.
47 kits in US and 48 kits in Sweden