Medical convenience kits Item Number/Description UICL44P ADULT CENTRAL LINE INSERTION SUPPLY KIT UICL44R ADULT CENTRAL LINE INSERTION SUPPLY KIT
Enforcement
- Recall Number
- Z-0177-2026
- Event ID
- 97659
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- American Contract Systems Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 22, 2025
- Initiation Date
- September 4, 2025
- Classification Date
- October 16, 2025
- Address
- 1601 Se Gateway Dr Ste 120, Grimes, IA, 50111-5223, United States
Description
Medical convenience kits Item Number/Description UICL44P ADULT CENTRAL LINE INSERTION SUPPLY KIT UICL44R ADULT CENTRAL LINE INSERTION SUPPLY KIT
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
UICL44P UDI-DI 191072194268 Lot 8003911 Bag serial number N/A; UICL44R UDI-DI 191072234018 Lot 8317711 Bag serial number 68244081 * If bag serial number is listed as "n/a", then entire lot is affected
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
32,433 total