Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.
Recall: Z-0176-2021
·
Reported October 28, 2020
Enforcement
- Recall Number
- Z-0176-2021
- Event ID
- 86419
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Bien-Air Surgery Sa
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 28, 2020
- Initiation Date
- September 11, 2020
- Classification Date
- October 21, 2020
- Address
- rue de l' ouest 2b, N/A, Le Noirmont, N/A, N/A, Switzerland
Description
Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.
Reason
The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.
Code Info
Model 1600407 / Catalog number 1600407-001 Serial Numbers: 16C0017 16F0013 16I0009 16K0017 17F0039 17I0015 17L0013
Distribution
US Nationwide distribution including in the states of CA, MI and MN.