FDA Enforcement Class II Terminated

Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288

Recall: Z-0175-2022 · Reported November 3, 2021

Enforcement

Recall Number
Z-0175-2022
Event ID
88748
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
AB SCIEX
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 3, 2021
Initiation Date
October 4, 2021
Classification Date
October 26, 2021
Termination Date
January 23, 2024
Address
71 Four Valley Dr, N/A, Concord, N/A, N/A, Canada

Description

Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288

Reason

The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making incorrect conclusions using incorrect results.

Code Info

Software version 3.4

Distribution

Worldwide distribution - US Nationwide distribution in the states of District of Columbia, Florida, Illinois, Louisiana, Maryland, Michigan, Missouri, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Utah, Wisconsin and the countries of Austria, Finland, Germany, Hong Kong, Italy, Japan, Norway, Spain, Switzerland, United Kingdom, Italy, France, and Japan.

Quantity

30