FDA Enforcement Class II Ongoing

Robotics Pack SAH, REF SAR530M

Recall: Z-0174-2024 · Reported November 1, 2023

Enforcement

Recall Number
Z-0174-2024
Event ID
93086
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
American Contract Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 1, 2023
Initiation Date
September 1, 2023
Classification Date
October 26, 2023
Address
7802 E Telecom Pkwy, N/A, Temple Terrace, FL, 33637-0928, United States

Description

Robotics Pack SAH, REF SAR530M

Reason

During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.

Code Info

UDI/DI00191072179678, Lot Numbers: 949231, 983231

Distribution

US Nationwide distribution in the states of FL, IA.

Quantity

96 cases