FDA Enforcement Class II Ongoing

BD Nexiva Closed IV Catheter System - Single Port REF 383516 20 GA X1.0o in (1.1 x 25 mm) 61 mL/min (3660 mL/hr)

Recall: Z-0174-2023 · Reported November 9, 2022

Enforcement

Recall Number
Z-0174-2023
Event ID
90972
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Becton Dickinson Infusion Therapy Systems Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 9, 2022
Initiation Date
October 12, 2022
Classification Date
November 3, 2022
Address
9450 S State St, N/A, Sandy, UT, 84070-3213, United States

Description

BD Nexiva Closed IV Catheter System - Single Port REF 383516 20 GA X1.0o in (1.1 x 25 mm) 61 mL/min (3660 mL/hr)

Reason

During needle withdrawal, the tip shield may prematurely separate from the catheter adapter and preventing the needle tip from being in the safely shielded (leaving the needle tip exposed).

Code Info

Catalog Number: 383516 UDI-DI Code: (01)00382903835164 Lot Number: 2168865

Distribution

U.S.: New York, Florida, Maryland, North Carolina, Virginia, Georgia, Rhode Island, New Jersey, Pennsylvania, Illinois, South Carolina, Maine, and Massachusetts. O.U.S.: N/A

Quantity

56,080 units