FDA Enforcement Class II Terminated

Dimension Vista Protein 1 Calibrator, PROT 1 CAL, Siemens Material Number (SMN) 10465664 (US) and 10469516 (OUS), Product Code KC710U (US) and KC710 (OUS); IVD

Recall: Z-0173-2018 · Reported December 6, 2017

Enforcement

Recall Number
Z-0173-2018
Event ID
77381
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 6, 2017
Initiation Date
May 23, 2017
Classification Date
November 28, 2017
Termination Date
March 30, 2018
Address
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States

Description

Dimension Vista Protein 1 Calibrator, PROT 1 CAL, Siemens Material Number (SMN) 10465664 (US) and 10469516 (OUS), Product Code KC710U (US) and KC710 (OUS); IVD

Reason

Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).

Code Info

Lot numbers: 6JC001, 6JC001C, 6JC001G (exp. date 2018-01-11); 6MC001, 6MC001C (exp. date 2018-04-29); 6CC001A, 6CC001B, 6CC001F (exp. date 2017-07-14); 6FC001A, 6FC001C, 6FC001E (exp. 2017-10-27); --- Lot # (UDI): 6CC001A (008427680244436CC001A20170714), 6CC001B (008427680244436CC001B20170714), 6CC001F (008427680244436CC001F20170714), 6FC001A (008427680244436FC001A20171027), 6FC001C (008427680244436FC001C20171027), 6FC001E (008427680244436FC001E20171027), 6JC001 (008427680244436JC00120180111), 6JC001C (008427680244436JC001C20180111), 6JC001G (008427680244436JC001G20180111), 6MC001 (008427680244436MC00120180429), 6MC001C (008427680244436MC001C20180429).

Distribution

Worldwide Distribution - US (nationwide) Internationally to Austria, Australia, Belgium, Canada, France, Germany, Great Britain, Italy, Japan, Netherlands, Portugal, Slovakia, Spain, Switzerland and Republic of Korea South

Quantity

N/A