FDA Enforcement Class II Ongoing

BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader.

Recall: Z-0172-2023 · Reported November 9, 2022

Enforcement

Recall Number
Z-0172-2023
Event ID
90866
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Becton, Dickinson and Company, BD Biosciences
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 9, 2022
Initiation Date
October 3, 2022
Classification Date
November 3, 2022
Address
2350 Qume Dr, N/A, San Jose, CA, 95131-1812, United States

Description

BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader.

Reason

Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.

Code Info

Corrected as of 11/04/2022: Catalog Number: 663028 UDI-DI Code: 00382906630285 Lot Numbers: 2031238 2063994 2095615 2145791 2221697 2237045

Distribution

U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay

Quantity

11,709 tubes (corrected 11/04/2022)