FDA Enforcement
Class II
Terminated
Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.
Recall: Z-0171-2020
·
Reported October 30, 2019
Enforcement
- Recall Number
- Z-0171-2020
- Event ID
- 83480
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- ICU Medical Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 30, 2019
- Initiation Date
- August 1, 2018
- Classification Date
- October 22, 2019
- Termination Date
- April 27, 2021
- Address
- 600 N Field Dr, Lake Forest, IL, 60045-4835, United States
Description
Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.
Reason
A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.
Code Info
All Plum 360 Infusers with software versions 15.10.00.010/.011/.012 and 15.11.00.017 when used with an ICU Medical MedNet CDL
Distribution
US nationwide distribution, and Canada
Quantity
47,152 devices