FDA Enforcement
Class II
Ongoing
Basic Biopsy Tray, Item Number NMBP44L
Recall: Z-0170-2026
·
Reported October 22, 2025
Enforcement
- Recall Number
- Z-0170-2026
- Event ID
- 97659
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- American Contract Systems Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 22, 2025
- Initiation Date
- September 4, 2025
- Classification Date
- October 16, 2025
- Address
- 1601 Se Gateway Dr Ste 120, Grimes, IA, 50111-5223, United States
Description
Basic Biopsy Tray, Item Number NMBP44L
Reason
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Code Info
UDI-DI 191072188120 Lot 8515411 Entire lot impacted
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Quantity
32,433 total