Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system
Enforcement
- Recall Number
- Z-0170-2024
- Event ID
- 93129
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Asensus Surgical Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 8, 2023
- Initiation Date
- September 15, 2023
- Classification Date
- October 30, 2023
- Termination Date
- September 26, 2024
- Address
- 1 Tw Alexander Dr Ste 160, N/A, Durham, NC, 27709, United States
Description
Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system
Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
UDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008, SN 00031, SN 00025, SN 00011, SN 00028, SN 00038, SN 00029, SN 00036, SN 00025, SN 00024, SN 00006, SN 00019, SN 00040, SN 00041, loaded with Software (SW) version 2.7.4
US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania
21 units