FDA Enforcement Class II Terminated

Syngo.plaza, Picture Archiving and Communications System(PACS), Model 10863171 Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific work flows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.

Recall: Z-0168-2018 · Reported November 29, 2017

Enforcement

Recall Number
Z-0168-2018
Event ID
77456
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 29, 2017
Initiation Date
May 22, 2017
Classification Date
November 21, 2017
Termination Date
July 2, 2018
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

Syngo.plaza, Picture Archiving and Communications System(PACS), Model 10863171 Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific work flows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.

Reason

Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.

Code Info

non-medical software application Check Tool V1 .0 (L TA Incomplete Archive Check Tool V1 .0)

Distribution

Distribution to MN

Quantity

2 units