FDA Enforcement
Class II
Terminated
COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological
Recall: Z-0168-2015
·
Reported November 12, 2014
Enforcement
- Recall Number
- Z-0168-2015
- Event ID
- 69583
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Iba Dosimetry Gmbh
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- November 12, 2014
- Initiation Date
- September 25, 2014
- Classification Date
- November 4, 2014
- Termination Date
- January 26, 2015
- Address
- Bahnhofstr. 5, N/A, Schwarzenbruck, N/A, N/A, Germany
Description
COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological
Reason
Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.
Code Info
N/A
Distribution
Worldwide Distribution-Distributed in the US including Guam and the countries of Germany, Austria, France, UK, Poland, Finland, The Netherlands, Spain, Turkey, Russia, and South Africa, Algeria, Mexico, Brazil, Venezuela, Columbia, Panama, China, Japan Philippines, Korea, India, Singapore, Thailand, Taiwan, Hong Kong, and Malaysia.
Quantity
124 units