Perkin Elmer Specimen Gate, Screening Center. Data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.

Recall: Z-0166-2016 · Reported November 4, 2015

Enforcement

Recall Number
Z-0166-2016
Event ID
72013
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Perkin Elmer Life Sciences Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 4, 2015
Initiation Date
August 24, 2015
Classification Date
October 26, 2015
Termination Date
January 21, 2021
Address
Laitun 4 Mustionkatu 6, N/A, Turku, N/A, N/A, Finland

Description

Perkin Elmer Specimen Gate, Screening Center. Data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.

Reason

It would be possible for an abnormal analytical test result value from an affected newborn to be reported with a default result interpretation of normal. The analytical test result value would be correct but the interpretation of the test result value would be incorrect.

Code Info

Software Version 1.6

Distribution

US distribution to FL, GA, and NV; and Canada

Quantity

6 units