FDA Enforcement Class II Terminated

Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity.

Recall: Z-0160-2015 · Reported November 12, 2014

Enforcement

Recall Number
Z-0160-2015
Event ID
58185
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 12, 2014
Initiation Date
March 1, 2011
Classification Date
October 31, 2014
Termination Date
November 6, 2014
Address
28 Howe St, N/A, Ashland, MA, 01721-1305, United States

Description

Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity.

Reason

Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line

Code Info

Catalog Number: 165720 Lot Numbers: MDI5, MD15B, MD22, ME19, MP13. Manufacturing dates March 15, 20 I 0 through December 13, 20 I 0. Catalog Number: 6375 Lot Numbers: MD15, ME19, MPI3 Manufacturing dates March 15, 2010 through December 13, 20 I 0

Distribution

Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.

Quantity

Catalog: 16579 ; Catalog 6375: 3256 units